water vapor transmission analyzer - Qinsun
Fabricante profesional de instrumentos de prueba
Línea directa nacional de consulta:+86 021-64208466

water vapor transmission analyzer

The water vapor transmission analyzer can measure the water vapor transmission rate of various mater

Correo electrónico: info@standard-groups.com

Contenido

image.png

The water vapor transmission analyzer can measure the water vapor transmission rate of various materials under different kinds of  restricted conditions. The test conditions can simulate the actual application environment. The test results are featured with accuracy and reproducibility.
 
Operating Principle

 Two independent gas streams are maintained on the two sides of a sample at the desired temperature. Humidity and gas flow rates are measured. The transmission rate through the sample is computed using mass balance.
 
 Operating Principle
 
Where

n = moles Φ = humidity M = flow rate
t = time p = equilibrium vapor pressure
ί = inlet o = outlet P = total pressure

Aplicación

The PMI Water Vapor Transmission Analyzer is capable of measuring water vapor transmission through porous media such as textiles, leathers, man made materials, membranes, nonwovens, and fabrics used in numerous high technology components and consumer products manufactured by a variety of industries.

The instrument has the unique ability to measure vapor transmission rate over a wide range of humidity, temperature, and pressure under gradients of humidity, temperature, and pressure encountered in application environments.
 
Características

· Humidity on any side can be maintained between 5 and 95%
· Any desired pressure gradient can be maintained.
· Any desired test temperature can be achieved.
· Simultaneous pressure and humidity gradients can be maintained
· Flat samples in a wide range of sizes can be accommodated
· Completely automated test implementation, data acquisition, data store and data simplification.
 
 Technical Parameters
 

Proyecto

Parámetros

Humidity (φ) measurement

Range: 5 - 95%
Accuracy: ± 2%

Differential pressure transducers

Range: 4 torr (2 mm Hg)
Accuracy: 0.015%

Mass Flow Transducers

Range: 5 L/min
Accuracy: 1%

Humidity (φ) control

Range: 0 - 100%
Accuracy: ± 1.5% (φ = 0.5) ± 5% (high & low φ)

Temperature

Range: RT - 100 °C
Control: ± 2%
Accuracy: 0.4 °C (low φ) - 0.8 °C (high φ) at 100 °C

Mass Flow Controller

Range: 2000 cc/min
Accuracy: 1%

Descargo de responsabilidad: El contenido anterior es para referencia y comunicación solo entre expertos de la industria, y no garantiza su exactitud o integridad. De acuerdo con las leyes y reglamentos relevantes y los reglamentos de este sitio web, las unidades o individuos que compren artículos relacionados deben obtener cualificaciones y condiciones de cualificación válidas.