Capillary Flow Porometer

Advanced Capillary Flow Porometers yield very objective, accurate and reproducible results, consider

Correo electrónico: info@standard-groups.com

Contenido

Advanced Capillary Flow Porometers yield very objective, accurate and reproducible results, considerably reduce test duration, and require minimal operator involvement. Advanced Porometers are fully automated and are designed for Linear turbulence-free test gas flow. The pressure is measured close to the sample and therefore, the correction term in the differential pressure measurement is minimized. Required amount of pressure is uniformly applied on the O-ring seals on the sample and the need for hand tightening the cap on the sample chamber to apply pressure on the O-ring is eliminated. Automatic addition of wetting liquid reduces test time appreciably. This sophisticated instrument has found applications in a wide variety of industries.
 
Operation Principle
 A wetting liquid is allowed to spontaneously fill the pores in the sample and a nonreacting gas is allowed to displace liquid from the pores. The gas pressure and flow rates through wet and dry samples are accurately measured. The gas pressure required to remove liquid from the pores and cause gas to flow is given by
D = 4 g cos q / p
where D is the pore diameter, g is the surface tension of liquid, q is the contact angle of liquid, and p is the differential gas pressure. From measured gas pressure and flow rates, the pore throat diameters, pore size distribution, and gas permeability are calculated.
 
Aplicación
 The Capillary Flow Porometer is used for R&D and quality control in industries worldwide such as filtration, nonwovens, pharmaceutical, biotechnology, healthcare, household, food, hygienic products, fuel cell, water purification, and battery. Samples often tested include filter media, membranes paper, powders, ceramics, battery separators and health care products.
 
Parámetros técnicos

ProyectoParámetros
Pore size range0.013-500μm
Permeability range1 x 10-10, 1 x 10-6 (microflow at cc/sec/m/torr)
Tamaño de la muestra0.5" - 2.5"diameter
Pressure range0-500psi
Pressurized gasClean, dry, compressed air or non-flammable, non-corrosive gases
Pressure transducer range0-500psi
Resolution1/60000
Precisión0.15% of reading
Mass flow transducer range10 cc/minute - 500 L/minute
Power requirement110/220 VAC, 50/60 Hz (Other power sources can be used )
Tamaño30" H x 19" W x 18.5" D
Peso100Ibs

image.png

Descargo de responsabilidad: El contenido anterior es para referencia y comunicación solo entre expertos de la industria, y no garantiza su exactitud o integridad. De acuerdo con las leyes y reglamentos relevantes y los reglamentos de este sitio web, las unidades o individuos que compren artículos relacionados deben obtener cualificaciones y condiciones de cualificación válidas.