- Evaluating the wear - resistance of hip joint prostheses: By simulating human movement states such
WhatsApp: +86 13816217984
Email: info@qinsun-lab.com
Standard
Parámetros | Especificaciones |
---|---|
Angular Motion Range | Flexion/Extension: ±100° Abduction/Adduction: ±30° Rotation: ±40° |
Max. Axial Force | ±4,500 N (1,000 N resolution) |
Load Control Accuracy | ±0.5% of reading (0–4,500 N) |
Frequency Range | 0.1–5 Hz (programmable waveforms) |
Temperature Control | 37±2°C (PT1000 sensors) |
Lubrication System | 15L capacity with protein skimmer and bubble trap |
Data Acquisition | 100 kHz sampling rate (16-bit resolution) |
Power Requirements | 220V/110V ±10%, 50/60Hz, 2,500W |
Dimensions (W×D×H) | 1,800 × 1,200 × 2,000 mm |
Peso | 950 kg (including bath and control cabinet) |
Noise Level | <65 dB (A-weighted) |
Descargo de responsabilidad: El contenido anterior es para referencia y comunicación solo entre expertos de la industria, y no garantiza su exactitud o integridad. De acuerdo con las leyes y reglamentos relevantes y los reglamentos de este sitio web, las unidades o individuos que compren artículos relacionados deben obtener cualificaciones y condiciones de cualificación válidas.