Mask Synthetic Blood Penetration Tester - Qinsun
Fabricante profesional de instrumentos de prueba
Línea directa nacional de consulta:+86 021-64208466

Mask Synthetic Blood Penetration Tester

Medical Mask Synthetic Blood Penetration TesterMask Synthetic Blood Penetration Tester is used for a

Correo electrónico: info@standard-groups.com

Contenido

Medical Mask Synthetic Blood Penetration Tester


Mask Synthetic Blood Penetration Tester is used for anti-synthetic blood penetration of medical surgical masks, and mainly used in the field of medical protection products. It can also be used for other textile coating material determination.Medical masks synthetic blood penetration tester is used to determine the resistance of medical protective masks to penetration by synthetic blood at different test pressures,and can also be used to determine the resistance to blood penetration of other coating materials.

Parameter:
Spray Distance:300mm+/-10mm
Spray Speed:450cm/s,550 cm/s,635 cm/s
Nozzle:Diameter: 0.8mm, Length::12.7mm
Spray time:0.01s-99.99s
Synthetic blood volume:2L
Power Supply:AC 220V  50Hz

Mask synthetic blood penetration tester Criterios:

GB 19083-2010 Technical requirements of medical face masks - Anti-penetration property of 5.5 synthetic blood.

YY/T 0691-2008 Clothing for protection against infectious agents-Medical face masks - Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected)  

YY 0469-2011 Technical requirements of medical face masks– Blood penetration test apparatus.

ISO 22609:2004 Clothing for protection against infectious agents - Medical face masks - Test method for resistance against penetration by synthetic blood (fixed volume, horizontally projected)  

ASTM F1862-07Standard Test Method for Resistance of Medical Face Masks to Penetration by Synthetic Blood (Horizontal Projection of Fixed Volume at a Know Velocity)

Face Mask Synthetic Blood Penetration Resistance Tester

Descargo de responsabilidad: El contenido anterior es para referencia y comunicación solo entre expertos de la industria, y no garantiza su exactitud o integridad. De acuerdo con las leyes y reglamentos relevantes y los reglamentos de este sitio web, las unidades o individuos que compren artículos relacionados deben obtener cualificaciones y condiciones de cualificación válidas.


Mensaje en línea

Detalles de contacto

Teléfono de la empresa

+86-21-6420 0566

Horas de trabajo

De lunes a viernes

Teléfono móvil:

13816217984

Correo electrónico:

info@qinsun-lab.com

Código QR