Dry Cleaning And Washing Cylinder - Qinsun
Fabricante profesional de instrumentos de prueba
Línea directa nacional de consulta:+86 021-64208466

Dry Cleaning And Washing Cylinder

Dry Cleaning And Washing CylinderTesting Standards:AATCC 162, 16 CFR Part 1610Applications:Dry Clean

Correo electrónico: info@standard-groups.com

Contenido
Dry Cleaning And Washing Cylinder

Criterios de prueba:
AATCC 162, 16 CFR Part 1610

Applications:
Dry Cleaning and Washing Cylinder is designed for determining color fastness to dry cleaning and chlorinated pool water. It also can be used to prepare samples for flammability testing.

Product Information:
Dry cleaning and washing cylinders are also known as dry cleaning tester and AATCC 162 tester. Dry cleaning and washing cylinder is used to test the color fastness of fabric to dry-cleaning and chlorine water washing. The Dry Cleaning And Washing Cylinder is comprised of stainless steel rotating test chamber with capacity of 7650ml, support frame, variable speed drive with digital timer.

Parameters:
1.Equipped with stainless steel rotating roller;
2.Height of tank: 330mm;
3.Diameter of in-tank: 220mm;
4.Rotating speed can be adjusted according to customer's requirements;
5.Overall dimension: 720×490×620mm(L×W×H);
6.Equipped with digital timer and displays;
7.Weight: 98kg.

Acción
Clean the dry cleaning cylinder
Prepare the hardness concentrates according to the relevant standards.
51 ml of the hardness concentrate was diluted to 5100 ml with distilled or deionized    water.
If necessary, adjust the pH of the solution to 7.0 with sodium carbonate or acetic acid.
5000 ml of the solution (made in the above 3) was charged into a dry-cleaning cylinder. The temperature was adjusted to 21 ° C.
Put the color test sample with the standard test fiber into the cylinder, turn off the barrel, and roll for 60 minutes.
Get out the sample and remove the excess solution.
 

Parámetros técnicos
ProyectoParámetros
Tamaño72 largo x 49 ancho x 62 alto (cm)
EstructuraAcero inoxidable
Tratamiento de superficieTratamiento de dibujo
Peso de transporte80 kg
Fuente de alimentación220V 50hz máximo 005A
 

Notes
1. For dry cleaning test, the instrument should be placed in the fume hood (using toxic solvent of tetrachloroethylene).
2. The operators need to wear gas masks and protective gloves.
3. Carefully handling with the liquid recycling, sending to the waste liquid recycling station for suggestion.

Exhibición de productos
 
 

Descargo de responsabilidad: El contenido anterior es para referencia y comunicación solo entre expertos de la industria, y no garantiza su exactitud o integridad. De acuerdo con las leyes y reglamentos relevantes y los reglamentos de este sitio web, las unidades o individuos que compren artículos relacionados deben obtener cualificaciones y condiciones de cualificación válidas.


Mensaje en línea

Detalles de contacto

Teléfono de la empresa

+86-21-6420 0566

Horas de trabajo

De lunes a viernes

Teléfono móvil:

13816217984

Correo electrónico:

info@qinsun-lab.com

Código QR